The University of Nottingham
Brain & Body Centre
Organization & Partners People Approaches & Tools Topics & Projects Publications Our Retreats Volunteers Users Families & Children Media Contact Us Home

Philips Achieva 1.5 Whole Body Fat Quantification - MR Acquisition Protocols - Users

Image 1

Philips Achieva 1.5 Whole Body Fat Quantification

Fat Quantification, Developed for a study of “Adiposity and cardiovascular and metabolic health” (PI: Pausova, Developer: John Totman)
Implemented at B&BC by John Totman (john.totman@nottingham.ac.uk)

Multi Station Scout

Dual Echo FFE
Sequence FFE Image 2
TR (ms) 382
TE (ms) First Echo 6.9
TE (ms) Second Echo 9.21
TFE Factor 20
Signal average 1
Orientation Transverse
Thickness (mm) 10
Gap (mm) 0
Number of slices 26
Field of View (mm) AP: 527 LR: 65%
Pixel Size (mm) Interpolated 1.5x1.49
Scan time (min) 0:27 per Dynamic
(Depends on Patient Height)
Dual Echo Water Selective FFE
Sequence FFE Image 3
TR (ms) 382
TE (ms) First Echo 6.9
TE (ms) Second Echo 9.21
TFE Factor 20
Fat Suppression Proset 1331 Water Selection
Signal average 1
Orientation Transverse
Thickness (mm) 10
Gap (mm) 0
Number of slices 26
Field of View (mm) AP: 527 LR: 65%
Pixel Size (mm) Interpolated 1.5x1.49
Scan time (min) 0:27 per Dynamic
(Depends on Patient Height)
Dual Echo Fat Selective FFE
Sequence FFE Image 4
TR (ms) 382
TE (ms) First Echo 6.9
TE (ms) Second Echo 9.21
TFE Factor 20
Fat Suppression Proset 1331 Fat Selection
Signal average 1
Orientation Transverse
Thickness (mm) 10
Gap (mm) 0
Number of slices 26
Field of View (mm) AP: 527 LR: 65%
Pixel Size (mm) Interpolated 1.5x1.49
Scan time (min) 0:27 per Dynamic
(Depends on Patient Height)